Home > Product Show

YZn-2310

YZn-2310 Automatic Online Monitor of Zinc in Water

In a certain pH environment, zinc reacts with a specific reagent to form a complex, with the maximum absorption at the specified wavelength. In 50 ml of solution, it contains 0 to 120 mg of zinc, which conforms to the Pear law. The complex is stable f…

Order Online
  • Product Principle
  • Product Features
  • Technical Parameters

In a certain pH environment, zinc reacts with a specific reagent to form a complex, with the maximum absorption at the specified wavelength. In 50 ml of solution, it contains 0 to 120 mg of zinc, which conforms to the Pear law. The complex is stable for several hours, and will fade slightly after being left for a long time.

Characteristics:

Valve components selection. It can select the reagent sampling sequence, with flexible channels, variable functions, minimum dead volume, easy maintenance, long service life and other advantages.

Printing function. It can be connected to a printer (optional order) to print the measured data in time.

It adopts 7-inch touch screen, with simple operation interface, easy for users to learn, operate, and maintain.

Super large data storage capacity. It can realize storage of historical data for no less than 5 years (measurement interval is 1 time/1h), which greatly meets the needs of customers.

Automatic leakage alarm function. When reagent leakage occurs, the instrument will automatically alarm to remind the user to perform maintenance.

Light signal recognition function to ensure high accuracy of quantification.

Simple and convenient chemical reagents maintenance. The reagents can be replaced once a month, which greatly reduces the maintenance of the user.

U disk one-click upgrade. The software update is much easier.

 

1 Method basis: spectrophotometry.

2 Measuring range: 0~10.0mg/L.

3 Accuracy: ±10% or ±0.001mg/L (the larger of the two).

4 Repeatability: 10% or 0.001mg/L.

5 Measurement period: The minimum measurement period is 30 minutes.

6 Sampling cycle: time interval (10~9999min arbitrarily adjustable) and hourly measurement mode.

7 Calibration cycle: adjustable at any interval any time from 1 to 99 days.

8 Maintenance cycle: 1 time/month.

9 Reagent consumption: less than 0.1 RMB/sample.

10 Display: 10-inch color screen, 1024*600, 16.77 million colors

11 Output: RS-232, 4-20mA (optional).

12 Communication protocol: Modbus RTU, new national standard protocol, dynamic control protocol, other

13 Environment requirements: indoor with adjustable temperature, recommended temperature +5~28℃; humidity ≤90% (no condensation).

14 Power supply: AC220±10% V, 50±10% Hz, 5A.